PULSER
Type
Negative square wave
Voltage
20 to 400 V, 1V steps.
Width
50 to 10000 ns, 1.0 ns steps
PULSER
Voltage
20 to 400 V, 1V steps.
Width
100 to 20000 ns, 5.0 ns steps
RECEIVER
Bandwidth
Low 22.5KHz to 20.5 MHz @-3dB
Mid Low 26.5KHz to 20.5MHz @-3dB
Mid High 108KHz to 20.5MHz @-3dB
High 390KHz to 20.5MHz @-3dB
Gain range
0 to 140 dB, 0.1 dB steps
TCG curve
80 dB maximum range, 16 points, 16 ms, 32 ns resolution, 80dB/us slope.
DIGITAL FUNCTIONS
A/D sampling rate
125 MSPS, 14 bits
Sampling decimation factor: 1 to 64
Gates
4 gates, max + 4 first echoes per gate.
1 gate for interface echo synchronization.
Signal modes
RF, True Envelope, Rectified (full, +/-).
Logarithmic full wave, Logarithmic envelope
Frequency Filters
Digital programmable IIR type, low-pass, high-pass, band-pass.
Noise reduction filters
Averaging
Anti-impulsive noise
Post-Rect smoothing filter
0 to 100% smoothing level control
Other digital functions
Real time alarms associated to echoes in gates.
Signal inversion.
Signal compression factor up to 64.
OTHERS
UT connectors
FISCHER D103A015
General I/O
4 digital inputs, 2 digital output (up to 24 V) optocoupled.
Fiber optic port
LC-Duplex optical connector (1 Gbit full-duplex)
Dimensions
FTT-2T: 175 x 115 x 64 mm (DxWxH).
FTT-2R: 175 x 115 x 64 mm (DxWxH).
Power
FTT-2T: 24 Vdc, 1.2 A max
FTT-2R: 24 Vdc, 0.6 A max
Accessories
FTT-2 PE Adapter: P/E Technique
GENIUS PUA-UT: High Sensibility Preamplifier 58dB Gain